5 Hygienic reprocessing
5.3 Hygienic reprocessing of the pressure
Requirement
Result

5.4 Cleaning and disinfection plan

Part
Cleaning
Using a lint-free, dry
Pressure reducer
or moistened cloth:
Use clean water.
22
EN
OXYWAY
reducer
Risk of injury if disinfectants containing alcohol or grease are
used for wipe disinfection!
Disinfectants containing alcohol or grease react with oxygen to
create a flammable mixture and can result in explosions and injury
to the user and patient.
 When wiping the pressure reducer with disinfectant, do not
use disinfectants containing alcohol or grease.
Risk of injury due to ingress of liquids!
Liquids entering the pressure reducer may cause explosions and
therefore injure the patient.
 Do not immerse the pressure reducer in liquids.
The pressure reducer is removed
(see "4.2 Removing the pressure reducer", page
1. Hygienically reprocess the pressure reducer in accordance with
the cleaning and disinfection plan
disinfection plan", page
2. Install the pressure reducer
reducer", page
3. Perform a function check
check", page
23).
The pressure reducer is hygienically prepared.
Disinfection
Wipe down with
Incidin™ OxyWipe S
(Ecolab). Treat all
surfaces at least twice
until they are visibly
clean.
(see "5.4 Cleaning and
22).
(see "4.1 Installing the pressure
15).
(see "6.2 Performing a function
Thermal
disinfection
Not permitted
16).
Sterilization
Not permitted