Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6.3 Repairing leaks

Requirement
The pressure reducer is installed
reducer", page
1. Prepare a soap solution using unperfumed soap.
2. Open the valve of the oxygen cylinder slowly.
3. Coat all the screw connections with the soap solution.
4. Dry the screw connections with a lint-free cloth.
5. Close the oxygen cylinder.
6. If the sealing ring has a leak: Replace the sealing ring
7. If the safety valve has a leak: Remove the pressure reducer
8. Check again for leaks.
Result
The leak has been rectified.
15).
Bubbles will form if a leak is present.
Replacing the sealing ring", page
"4.2 Removing the pressure reducer", page 16)
repaired.
6 Function check
(see "4.1 Installing the pressure
29).
OXYWAY
(see "8.5
(see
and have it
EN
25

Publicité

Chapitres

Table des Matières
loading

Produits Connexes pour Weinmann OXYWAY Fix I

Table des Matières