Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20

7 Filter Capacity

Capacity Tables
HYDROS Water Treatment
System
Total hardness
Volume
in °GH
in litres
4
12.500
5
10.000
6
8.333
7
7.143
8
6.250
9
5.556
10
5.000
11
4.545
12
4.167
13
3.846
14
3.571
15
3.333
16
3.125
17
2.941
18
2.778
19
2.632
20
2.500
21
2.381
23
2.174
25
2.000
28
1.786
31
1.613
35
1.429
Note: The stated capacities are guide values that may vary by ± 20% depending on the
product volume flow, local water quality and machine type.
8 Repair
Regularly check the filter system for leaks. Regularly check the hoses for kinks. Bent
hoses must be replaced.
The complete filter system must be replaced in rotation after a maximum of ten years.
The hoses must be replaced in rotation after a maximum of five years.
Caution: Prior to exchange read the technical data (Chapter 10) and the operating and
!
safety information (Chapter 3).
Regularly clean the outside of the filter system with a soft, damp cloth.
Caution: Do not use any substances incompatible with the material (Chapter 3.4) or
!
astringent, abrasive cleaning agents.
PURITY 1200 Clean Extra
Total hardness
Volume
in °GH
in litres
4
12.500
5
10.000
6
8.333
7
7.143
8
6.250
9
5.556
10
5.000
11
4.545
12
4.167
13
3.846
14
3.571
15
3.333
16
3.125
17
2.941
18
2.778
19
2.632
20
2.500
21
2.381
23
2.174
25
2.000
28
1.786
31
1.613
35
1.429
Volume
in litres
13.611
10.889
9.074
7.778
6.806
6.049
5.444
4.949
4.537
4.188
3.889
3.630
3.403
3.203
3.025
2.865
2.722
2.593
2.367
2.178
1.944
1.756
1.556
17

Publicité

Table des Matières
loading

Produits Connexes pour SciCan HYDROS

Table des Matières