6.1 Removing the Filter Cartridge
Switch off the power supply of the HYDRIM unit (remove plug).
Close the inlet valve
Place the flush hose in a bucket and remove pressure from the filter system by
opening the flush valve
Note: If the escaping water is more than 1 litre, the inlet value
closed or is blocked with scale.
Stand on the kick loops
vessel lid
as far as it will go.
Place the pressure vessel lid
Stand on the kick loops
vessel
Take your feet off the kick loops
hands on the mantle handles
Remove used filter cartridge
Place the used filter cartridge
drainage (> 5 min).
Lock the used filter cartridge
cartridge and discard according to the local regulations.
6.2 Inserting the Filter Cartridge
16
at the inlet hose
6
. Collect waste water in the 10L bucket.
15
by pressing the lock
3
counter-clockwise by the mantle handles
1
and press the pressure vessel
9
.
11
from the pressure vessel
2
in the sink with the connection facing down for
2
with the transport protection cap
2
Check the O-ring seal of the new filter cartridge
the groove, dirt and damage. Note: The cartridge seat has been
lubricated with food-safe lubricant at the factory.
Check the connector seat of the filter cartridge O-ring
pressure vessel lid
Place new filter cartridge
Stand on the kick loops
whole turning clockwise until the mantle handles
loops
.
9
Stand on the kick loop
vessel lid
on the pressure vessel
3
marking on the lid handle
Press the pressure vessel lid
lock engages
.
8
Switch on electrical power supply to the HYDRIM unit.
Execute the steps described under 5.3 to flush and bleed the new
filter cartridge
.
2
.
5
with both feet while opening the pressure
9
and turning it counter-clockwise
8
vertically on the lid handle
3
with both feet while turning the pressure
9
for dirt and damage.
3
in the pressure vessel
2
with both feet; lift the pressure vessel
9
with both feet and place the pressure
9
. The positioning of the arrow
1
must line up with the "INSERT" groove.
13
down and turn clockwise until the
3
is not completely
6
.
13
as far as it will go.
11
down with both
1
.
1
of the new filter
12
for correct seat in
2
in the
2
.
1
are over the kick
11
1