Mode Selection

4
Figure 4
Note: In this text, the international graphic symbol:
will be written as |O|O|.
Note:
Intermittent timing is preset for 8 seconds "OFF" and 15 seconds "ON" , starting
in the "OFF" cycle. Timing is preset at the factory using a source vacuum level of 500
mmHg (66.7 kPa). Timing may vary in relation to the source vacuum pressure.
|O|O| (INTER) - Suction is intermittent (cycled "on" and "off") and the suction
level can be adjusted with the suction control knob.
O (OFF) - No suction is supplied to the patient.
| (CONT) - Suction is continuous and can be adjusted with the suction control
knob.
5
Figure 5
Setting the suction level
1. Clamp the connective tubing to occlude the fitting port
Clamp
0
(INTER)
0
0
0
(OFF)
(INTER)
0
0
(INTER)
(CONT)
0
0
(OFF)
0
(OFF)
(CONT)
(CONT)
0
(INTER)
0
0
0
(OFF)
(INTER)
0
0
(INTER)
(CONT)
0
0
(OFF)
0
(OFF)
(CONT)
(CONT)
0
(INTER)
0
0
0
(OFF)
(INTER)
0
0
(INTER)
(CONT)
0
0
(OFF)
0
(OFF)
(CONT)
(CONT)
6
6700-0359-000
Intermittierende
Absaugvorrichtung
100
120
14
12
140
16
18
20
160
22
24
26
180
0
(INTER)
0
0
(OFF)
(CONT)
(Rev. 9) 08/2020
80
60
10
8
40
6
4
2
20

Publicité

Table des Matières
loading

Table des Matières