DISS Fitting
1. Insert the safety trap into the regulator fitting. Situate the tubing in the
desired position.
2. Turn the DISS wing nut clockwise to engage threads and tighten (there is
no O-ring, so the vacuum seal depends on a tight connection).
 Regulator
 Wing nut
 Safety Trap

Mode Selection

2-Mode Continuous
| (ON) - Suction can be adjusted with the suction control
O (OFF) - No suction supplied to the patient
3-Mode Continuous
MAX - Maximum, full-line vacuum is supplied to the patient
Note: Available on Three-Mode Vacuum Regulators ONLY
Setting the Suction Level
1. Move the mode selector switch to | (ON).
2. Push and rotate the suction control knob until the vacuum gauge indicates the required setting.
g
CAUTION: The suction control knob must be completely pushed in to adjust the vacuum level. Failure to
do so may damage the vacuum regulator.
8
Operation
8700-0007-000 (Rev. 16) 10/2020