10. Turn the mode selector knob to |O|O|.
11. Clamp the connective tubing.
12. Ensure that the timing cycles are 15 second "On" and 8 seconds "OFF" both
with a tolerance of ± 3 seconds by observing the gauge needle.
Note: The ISU starts in the off cycle.
13. Reduce the suction level to zero and turn the mode selector to O (OFF).

Patient Setup

1. Make sure the Pre-Use Checkout Procedure has been performed.
8
Figure 8
2. Turn the mode selector switch to | (CONT) and clamp tubing.
Clamp
3. Set the prescribed suction level.
WARNING
⚠ The regulator must be occluded when setting the prescribed
4. Turn the mode selector switch to O (OFF).
5. Attach the patient tubing to the collection canister or liner's patient port.
6. Turn the mode selector knob to | (CONT) or |O|O| (INTER).
suction level so that the patient does not receive higher than
required suction.
8
400
300
500
50
60
40
200
High Flow
70
30
High Vacuum
600
80
20
100
90
10
100
700
0
(INTER)
0
0
(OFF)
(CONT)
Underwater Seal Line
6700-0359-000
(Rev. 9) 08/2020
2 cm H
O
2
Made in USA