The bar graph in the display and the scale on the unit indicate
the range of the blood pressure which has been recorded.
If the values for systolic and diastolic pressure are in two dif-
ferent WHO ranges (e.g. systolic in the high-normal range and
diastolic pressure in the normal range) the graphic WHO clas-
sification on the unit indicates the higher range (high-normal in
the example described).

6. Saving and displaying measurements

• The results of each successful measurement are stored to-
gether with date and time. With more than 60 items of meas-
ured data, the earliest items of data measured are lost.
• Select the desired user memory by pressing the
and then the
button. Pressing the +/M button displays the
average value of the 3 most recent measurements stored in
the user memory. If you continue pressing the +/M button, the
latest individual results are displayed with date and time.

7. Deleting stored values

To delete individual measurements, you must first select them.
Press and hold the memory button +/M until CL appears in the
display. Then confirm the deletion with the user selection button
(CL and 00 appear in the display).
To delete measurements from a user memory, you must first
select them. Press the memory button +/M and the average
value of the last 3 saved measurements is displayed. Then
press and hold the memory button +/M again until CL appears
in the display. Confirm the deletion with the user selection but-
ton
(CL and 00 appear in the display).

8. Error messages / troubleshooting

In case of faults, the E_ message appears in the display.
Error messages can occur when
1 the cuff is too tight or too loose (Er1),
2. you move or talk during the measurement (in addition to (E2),
the cardiac arrhythmia icon
3. the cuff is inflated for longer than 25 seconds (E3).
4. errors occur during measuring (E4),
5. inflation pressure is higher than 300 mmHg (E5).
6. the batteries are almost empty
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
button
not move or talk. Re-insert the batteries if necessary, or else
replace them.

9. Cleaning and Care

• Clean your blood pressure computer carefully using only a
slightly moistened cloth.
• Do not use any detergents or solvents.
• Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.

10. Technical details

Model no.
Measurement
method
Measurement range
16
appears in the display),
(E6).
SBC 28
Oscillating, non-invasive blood
pressure measurement on the wrist
systolic 50 – 250 mmHg,
diastolic 30 – 200 mmHg,
pulse 40 –180 beats/minute