Automatic deactivation settings
When automatic deactivation is selected, the device detects the end
of coagulation and deactivates high frequency mode.
Press function key
2.
The display shows:
Note
After a waiting time of 30 seconds, or when the user presses function
key
, the system returns to the main menu.
Press function key
5.
The menu Special Functions is displayed:
Set automatic deactivation
shown on the display.
Confirm the setting by pressing function key
main menu.
- or -
To exit the menu without adopting the changes:
Press the function key
The coagulation gradient for deactivation can be adjusted in two steps
and applies equally to all monopolar and bipolar outputs:
• Long: A higher coagulation gradient is reached until deactivation
(forced coagulation)
• Short: A lower coagulation gradient is reached until deactivation
(soft coagulation)
If the function autoCOAG on/off is selected, the maximum dosage is
limited to 50. At the Long setting, the deactivation behavior changes
for the bipolar outputs. The HF power is activated as soon as the for-
ceps touches tissue; it is deactivated as soon as the forceps does not
touch any tissue anymore.
Note
The higher the dosage setting, the quicker the deactivation threshold
will be reached, and the more superficial is the coagulation.
48
High frequency surgical device
, using function key
and return to the
or wait for 30 seconds.
The setting autoCOAG on/off should not be used in
endoscopic applications, since there would be a risk
of accidental activation, e.g. when the instrument
WARNING
is introduced through the trocar.
Selecting the argon function
The argon control unit is connected to the HF device via the CAN bus
interface 18. Activation signals at the handle or foot pedal are trans-
mitted to the argon control unit.
Press function key
The display shows:
Note
After a waiting time of 30 seconds, or when the user presses function
key
, the system returns to the main menu.
Press function key
The menu Special Functions is displayed:
, as
To select the argon function
shown on the display.
Communication between the HF device and the argon control unit:
Communication correct:
Communication interrupted:
Confirm the setting by pressing function key
main menu.
- or -
To exit the menu without adopting the changes:
Press the function key
A monopolar output is assigned to the argon function. The following
settings are available:
• OFF: The argon function is deactivated
• Monopolar 1 (argon function at output 9)
• Monopolar 2 (argon function at output 10)
(GN 640)
2.
5.
, press function key
Display
Display
and return to the
or wait for 30 seconds.
as