Télécharger Imprimer la page

Agilent Technologies TwisTorr 84 FS Notice De Mode D'emploi page 324

Masquer les pouces Voir aussi pour TwisTorr 84 FS:

Publicité

17
Technical Information
Connection of the High Vacuum Flange
Attach the units and tighten each screw in turn. Repeat the
sequential tightening until the flange faces meet. Tighten the screws
with a torque of 20 Nm.
For ConFlat flange connections we recommend using Agilent
hardware.
To facilitate assembly and dismantling, apply Felpro C-100 high
temperature lubricant to the screw threads protruding from the
flange and between the nuts and flange.
Note that, in some cases, the connections can be made only with the
bolt in the lower side.
Exercise care when tightening nuts and bolts to avoid creating dents in the
CAUTION!
envelope as this may cause the pump rotor to lock.
324/336
TwisTorr 84 FS User Manual / 87-901-043-01(B)

Publicité

loading

Ce manuel est également adapté pour:

X3502-64000X3502-64001X3502-64002X3502-64003