16
Instructions for Use

Pump Fixing

Pump Fixing

WARNING!
Failure to comply with these installation instructions could result in the pump
detaching from the system in the event of a rotor failure, which could cause
property damage or serious injury or death.
The TwisTorr 84 FS can be installed in any position. Fix the TwisTorr
84 FS in a stable position, mounting the inlet flange of the turbopump
to the system counter-flange, with a connection capable of
withstanding a torque of 250 Nm around its axis.
The ISO-K flange can be fixed using high strength steel clamps (like
Agilent model IC63250DCMZ).
The following table shows, for each flange, the necessary number of
IC63250DCMZ clamps and the relevant fixing torque.
FLANGE
ISO 63 K
The turbopump with ConFlat inlet flange must be fixed to the
vacuum chamber by means of the appropriate Agilent hardware. See
the appendix "Technical Information" for a detailed description.
The TwisTorr 84 FS cannot be fixed by means of its base.
NOTE
272/336
FIXING DEVICE
M10 clamps
TwisTorr 84 FS User Manual / 87-901-043-01(B)
N.
FIXING TORQUE
4
22 Nm