to store its last settings. If the "History" parameter is turned off, all parameters and options will return to their default
settings when you leave that application. (Unless it has been saved as a method).
Applications
Single wavelength
This makes simple absorbance (A) and % transmission (%T) measurements on samples, measuring the amount of light
that has passed through a sample relative to a reference (this can be air).
The procedure is as follows:
Step 1
Set wavelength by using keypad numbers or left and
right arrows.
Press the down arrow key.
Step 2
Select the mode, Absorbance or %T, using the left and
right arrows.
Press the down arrow key.
Step 3
Select whether or not to adjust the name of your
sample before each new run, using left and arrow keys
to select Yes or No.
Step 4
To enter the results screen with the selected
parameters press OK
and return to the Applications Folder by pressing
Cancel
.
Step 5
Insert the reference. Press the reference key
This will be used for all subsequent samples until
changed.
Step 6
Insert sample and press
Repeat step 6 for all samples.
or Cancel the selections
.
.