GENDEX 765DC Manuel De L'opérateur page 17

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Leakage Radiation:
X-Ray Coverage:
1
AHE Aluminum Half value Equivalent correlation
2
SSD Source to Skin Distance
Weight:
Maximum Head Voltage:
Thermal Characteristics:
Housing Heat Storage Capacity
(see Graph for more detail)
Maximum Cooling Rate
45
40
35
30
25
20
Complies to 601-1-3 IEC Section 29.204 "Leakage
Radiation". Maximum continuous power input equals
1,400 heat units per min. (HU/Min.) tube is the limiting
factor. HU/Min. = (kV) x (mA) x (1.4) x (Exp. Time
(sec) / 60)
1,400 HU/Min. = 65 kV x 0.27mA x 1.4 x 60 sec.
The insert target angle is 12.5°
13.0 lbs.
Anode to Cathode
Anode to Ground
Heat Units = (kV * mA * time in seconds * 1.4)
Gendex 765DC Housing
Heating & Coling Chart
Time (min.)
Printed on: 29 Mär 2012, 05:33:23 pm; Printed by: UWE.ZELLER
Circular cone at 8 inches (200 mm) SSD
Dia. +0, -3mm.
Circular cone at 12 inches (300 mm) SSD
Dia. +0, -3mm.
Rectangular cone at 8 inches (200 mm) SSD
x 46mm
5.9 kg.
250 kHU's (177J)
3.3 kHU's/Min.
Heating, 100% Duty Cycle, ambient
temperature 20°C
Cooling, ambient
temperature 20°C
765DC
2
60 mm
2
60 mm
2
36mm
65 kV
70 kV
300
250
200
150
100
50
0