Changing the Digital Option
The 765DC is set at the factory for use with the Gendex DenOptix digital imaging
system. This can be confirmed by turning the system on and pressing the button to select
digital. The time displayed should now indicate 0.160 seconds.
1. Turn the system off.
2. Press and hold the anatomical time selector switch while turning the system on (see Figure 1).
3. The time display will now read "SP 1,2, or 3" depending on which film type is configured.
4. Press the film or digital selector switch until the display reads "D1". Press the exposure time
'UP' button. The display will now indicate "D2". This indicates that the system is now
configured for a CCD type digital imaging system.
5. Press the anatomical time selector switch again to store the new setting.
Changing the power on selection
The 765DC is set at the factory to have the film option for an adult selected when the
unit is first turned on. If desired, the unit can be set so that it automatically selects the
digital option or a child instead when the control is turned on.
1. Turn the system off.
2. Press and hold the anatomical time selector switch while turning the system on (see Figure 1).
3. The time display will now read either "SP 1,2, or 3" or "D1" or "D2" depending on
whether the film or digital option is selected. Note that the green indicator light next to the
selection is flashing.
4. To select for the unit to come on in the digital mode, make certain that the green light is
flashing next to the digital option. Press the anatomical selector switch to store the
information and turn the system off. The unit should now automatically select the digital
option when first turned on.
5. To select for the unit to come on in the child mode, press the adult or child selection button
until the green light is flashing next to the child option. Press the anatomical selector switch
to store the information and turn the unit off. The unit should now automatically select the
child option when the unit is first turned on.
Modifying the system for use with the optional 12" (30 cm) cone
The 765DC is set at the factory for use with the standard supplied 8" (20 cm ) cone.
The 12" (30 cm) long cone is recommended when using the paralleling film positioning
technique along with the Rinn XCP kit. Using the long cone requires the use of longer
exposure times, which can be programmed into the system by changing an internal
electrical setting.
This modification should be performed only by a qualified Gendex service
agent. Contact your Gendex service agent to make the change.