VI.3 - Lubrication
Follow the spray lubricant manufacturer's instructions.
Wear appropriate protective clothing (splashes, etc.).
E
Keep away from all sources of heat or ignition.
In particular: do not smoke. Risk of infl ammation.
After cleaning and before each sterilisation (once a day minimum):
• Remove the rotary instrument,
• Insert the tip to the back of the MD (Fig 8),
• Cover the head of the MD with a soft cloth or paper or a wipe.
• Point the head towards the bottom,
• Spray several times by fi rmly holding the instrument,
• Wipe excess oil with a cloth or a wipe.
VI.4 - Operational test
• Connect the MD to a micro-motor, point the head towards the bottom.
• Turn the micro-motor on at low speed for 30s. Gradually bring up to
maximum speed.
• Wipe excess oil with a cloth or a wipe
If you notice overheating, irregularities, vibrations, abnormal noises
x
when using the MD, immediately contact your after-sales service.
14
VI.5 - Sterilisation
• The instruments are to be sterilised before fi rst use and after each use.
• Only sterilise cleaned, lubricated and tested instruments.
• We recommend sterilisation that complies with the EN 13060 standard,
class B. All other methods are forbidden.
• Independently put each MD in a sterilisation sachet that complies with
current standards and the size of the MD.
• Comply with the space between the sachets and do not overload the
autoclave.
• These MDs and their accessories (apart from accessories with the
g
logo
W
which are single use) must be sterilised at 135°C for 18 minutes
minimum (sterilisation time).
x
• Check that the device is completely dry at the end of the cycle.
VI.6 - Storage
D
Keep the MD in a sterilisation sachet in a clean and dry place.