CombiLyz AFI4/AFI5
Inductive conductivity transmitter
Example of mounting with weld-in sleeve ZPW2-521
6.

Approvals

The EHEDG certificate is only valid in connection with the appropriate installation parts.
These are marked with the "EHEDG Certified" logo.
The 3-A Sanitary Standard requirements are only met with the appropriate installation parts.
These are marked with the 3-A logo.
Approved by Underwriter Laboratories (UL) for use in the USA and Canada as an industrial
control device.
For more information on approvals and certification, refer to the product page on www.baumer.com.
7.

Electrical connection

7.1 External connections

9 A voltage supply of 15 to 35 V DC is provided.
Switch off supply voltage.
f
Connect the sensor in accordance with pin assignment.
f
NOTICE
Damage to sealing or connector!
Tightening the knurl with too much force can damage the connector or the O-ring inside the cable gland.
Only tighten the knurl by hand with a maximum torque of 0.6 N.
f
Do not use any tools when tightening the knurl.
f
8 / 52
Conductivity measurement
1
1
1
2
Operating Instructions
1
ZPW2-521
2
Leakage hole
www.baumer.com