10. Troubleshooting

Fault
Display is off and no
signals from transmitter
Display is on but no signals
from transmitter
Display is off but there are
signals from transmitter
Display does not show the
correct data

11. Cleaning, maintenance and repair

Cleaning
Clean, disinfect or sterilize sensor as needed
f
(CIP/SIP).
Maintenance
Regular maintenance is not required.

12. Disposal

Do not dispose of in household
f
waste.
Separate materials and dispose
f
of in compliance with nationally
applicable regulations.

13. Accessories

For adapter and other accessories, refer to www.baumer.com.

14. Technical data

Conductivity performance characteristics
Conductivity
Min. measurable con-
ductivity
Max. measuring span
Min. measuring span
14 / 52
Conductivity measurement
Cause
Sensor not correctly connected
Device error
Short circuit
UnitCom cable not connected
Unsuitable media characteristics
14 selectable ranges
50 µS/cm
1000 mS/cm
500 µS/cm
Operating Instructions
Action
Check plug and power supply.
f
Dismount and return sensor.
f
Remedy short circuit.
f
Connect the UnitCom cable between
f
display and transmitter.
Check signal quality with
f
FlexProgrammer 9701.
Repair
Do not repair the sensor yourself.
Send damaged sensor to Baumer.
f
Max. measuring error
Reference conditions for
max. measuring error
± 1.0 % FSR,
0 ... 1 mS/cm
to 0 ... 500 mS/cm
± 1.5 % FSR,
0 ... 1000 mS/cm,
± 1.5 % FSR,
0 ... 500 µS/cm
Sensor incl. transmitter
at 25 °C ambient tempe-
rature