1. Press SET until the display shows [SEt rEF] and press CAL to enter, with Ù or
Ú, the reference type of the redox electrode [CALOMEL] or [AgCL]. Press CAL
to continue or SET to return.
2. The display shows e.g. [rEF=25°C]. Select the temperature to which all future
conductivity measurements will be referred with Ù or Ú and press CAL to con-
tinue.
1. Press SET until the display shows [SEt dAtE] and press CAL to continue.
2. The display shows [dAtE=on] or [dAtE=OFF]. Select the desired mode with Ù or
Ú and press CAL to continue.
3. Program the date with Ù or Ú (respectively year, month and day) by pressing
CAL each time.
1. Press SET until the display shows [SEt tiME] and press CAL to continue.
2. The display shows [tiME=on] or [tiME=OFF]. Select the desired mode with Ù or
Ú and press CAL to continue.
3. Program the time with Ù or Ú (respectively hour, minutes and seconds) by
pressing CAL each time.
1. Press SET until the display shows [SEt °C] and press CAL to continue.
2. The display will show the measured temperature e.g. [°C=23.7][Aut]. If neces-
sary, calibrate your probe with Ù or Ú and press CAL to continue or SET to
return. Press RES to reset the temperature calibration to the original factory
calibration.
3. When no Pt1000 temperature probe is connected to the instrument the display
will show the manual temperature e.g. [°C=23.7][MAn]. Shift to the proper
temperature with Ù or Ú and press CAL to continue.
1. Press SET until the display shows [SEt rEC] and press CAL to continue.
2. The display shows for all channels e.g. [rEc=PH], [rEc=MV], [rEc=EC],
[rEc=°C]...etc. Select the desired mode with Ù or Ú. Press CAL to continue.
3. This step is skipped unless [rEc=EC] has been chosen. Press CAL and the display
shows the conductivity measuring range & units, e.g. [rEc 0.01MS][Unit]. Select
the desired range with Ù or Ú and press CAL to continue.
4. The display shows e.g. [rEc=6.00][LoW]. Select the desired minimum level cor-
responding to a 0 V recorder output with Ù or Ú and press CAL to continue.
5. The display shows e.g. [rEc=8.00][HiGH]. Select the desired maximum level cor-
responding to a 2 V recorder output with Ù or Ú and press CAL to continue.
C732
C732
Reference
Date
Time
Temperature
Recorder
Recorder