3. Dispensing procedure
a) For safety make sure that the discharge tube does not point at the user or any other
person. By using the built-in FIX adapter it is possible to fit the dispenser accordingly.
Remove stopper cap by holding the discharge tube tight with other hand.
b) Set the desired volume with the volume setting system.
c) Lift the dispensing piston up with the finger guard. Do not touch the piston itself to
avoid contact with chemicals. The piston stops at the volume set. The graduation on
the glass cylinder is an additional optical control of the volume set or air-bubbles.
d) To dispense the liquid push the piston down.
e) Avoid uneven piston movements or hard pushes against the stopper ring.
f) Replace stopper cap after the dispensing procedure by holding the discharge tube
tight with other hand.
Warning:
The piston must be pushed down manually. Therefore chemicals may remain
in the dispensing cylinder after dispensing. To avoid unintentional dispensing,
always dispense the total volume set. The stopper cap must be placed on the
discharge tube after every use.
When dispensing observe safety precautions!
V. Operating Temperature
The operating temperature is from +15° C (+59° F) to +50° C (+1° F).
VI. Sterilization
The dispenser can be autoclaved at +11° C (+48° F) / 1 bar. The piston can remain
in the cylinder when autoclaving.
Before sterilization loosen screw of the volume setting system.
Warning: Do not autoclave with higher temperatures. This will damage the
dispenser. Wait before use until the dispenser has room temperature.
VII. Transport and Storage
1. When transporting the dispenser never carry on the finger guard but always
on the PP valve block to avoid unintentional dispensing of liquid! Replace
the stopper cap before transporting and fix the volume setting system at the
zero-position.
. Please transport and store the dispenser only in the original box to avoid damages.
Storage temperature: -10° C (+14° F) to +50° C (+1° F).