NOTE: the current convention arrow on the probe to get the proper polarity
reading.
NOTE: clamping around both the "hot" and "neutral" wires may give a zero
reading.
To increase the measurement sensitivity of the current probes, loop additional
turns of the wire under test through the jaws. See Figure 4. The sensitivity of the
current probe is multiplied times the number of loops in the jaws. For example:
200 mV/A X 4 turns = 800 mV/A.
Figure 4: Increasing the sensitivity
Remember to unclamp the probe from the conductor before disconnecting it from
your meter or instrument.