Drive SmartDose EVOLUTION OM-975 Mode D'emploi page 19

Publicité

NOTE: Ensure that the oxygen inlet remains free of dust, etc.
STEP 3: Disinfection Instructions:
Saturate a clean, lint-free cloth with a suitable disinfecting agent (e.g., 1:5 chlorine bleach [5.25%] and water solution) and wipe the
exterior of the device.
NOTE: Ensure that the device remains visibly wet for ten minutes, with the solution reaching all crevices and hard-to-reach areas.
WARNING: Do not wet the area of the oxygen inlet connection, the disinfectant residue may create a risk of fire. Do not submerge the
conserver in the disinfectant, wipe disinfect only.
STEP 4: Allow all parts to dry completely before checking for proper operation.
STEP 5: Check the oxygen conserver for proper operation:
• Press and hold the Selector Button and ensure that each pulse setting can be selected and that each Dosage LED illuminates.
• Attach the device.
• Ensure the pressure gauge is working.
• Ensure the flow is working.
• Set the device to each setting and verify that oxygen pulses.
If the device is operating properly, it is ready to be used by another patient. If the device is not operating properly, the
provider should return the device to Drive Medical for service.
NOTE: There is no portion of the gas pathways through the conserver that could be contaminated with body fluids or expired gas under
normal or single fault conditions.
PART/ACCESSORY
Outside surface of the conserver and controls
Oxygen Tubing, nasal cannula
WARNING: Do not attempt to open or remove enclosure. There are no user-serviceable internal components.
NOTE: The disinfection process can only be performed by an appropriately trained individual.
RECOMMENDED DISINFECTION INTERVAL
Between patients
Do not clean, replace between patients
COMPATIBLE DISINFECTION METHOD
1:5 Chlorine Bleach (5.25%) and water solution
N/A
19