6 Function check
6 Function check
6.1 Intervals
6.2 Performing a function check
Result
46
EN
prismaCONNECT
Perform a function check at regular intervals:
•
After every hygiene treatment
•
After every repair
•
At least every 6 months
1. Check whether the module has been set up and
connected properly in accordance with the instructions
for use.
2. Check housing for cracks, damage, and severe soiling.
3. Connect the module to the device
the prismaCONNECT module", page
4. Check whether the symbol
shown in the display of the therapy device.
5. Connect the Ethernet cable to the LAN port and to a PC or
an IT network.
6. Check whether the network symbol
display of the therapy device.
7. If the module is not working properly or is damaged:
Contact your specialist dealer.
The function check is complete.
(see "4.1 Connecting
39).
for prismaCONNECT is
is shown in the