Scope Of Supply; Specification And Ordering Information - Camille Bauer SINEAX B 811 Mode D'emploi

Source d'alimentation
Masquer les pouces Voir aussi pour SINEAX B 811:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operating Instructions
Power pack SINEAX B 811
Contents
1. Read first and then... ................................................ 23

2. Scope of supply ........................................................ 23

4. Brief description ........................................................ 24
5. Technical data ........................................................... 24
6. Overview of the parts ................................................ 26
7. Exchanging frontplates ............................................. 26
8. Withdrawing and inserting the device ....................... 26
9. Mounting ................................................................... 26
10. Electrical connections ............................................... 27
11. Configuration ............................................................ 31
12. Commissioning ......................................................... 31
13. Releasing the power pack ........................................ 32
14. Dimensional drawings ............................................... 32
1. Read first and then ...
The proper and safe operation of the device
assumes that the Operating Instructions are
read and the safety warnings given in the
various Sections
9. Mounting
10. Electrical connections
12. Commissioning
are observed.
The device should only be handled by appropriately
trained personnel who are familiar with it and authorised
to work in electrical installations.
2. Scope of supply
(Fig. 1)
(5)
(1)
Betriebsanleitung
Speiseferät SINEAX B 811
Mode d'emploi
Source d'alimentation SINEAX B 811
Operating Instructions
Power pack SINEAX B 811
B 811 B d-f-e
999 469
Aargauerstrasse 7
CH-5610 Wohlen/Switzerland
Telefon +41 56 618 21 11
Telefax +41 56 618 24 58
Camille Bauer AG
Telex 827 901 cbm ch
Power pack (1)
2 withdrawing handle (2) (for withdrawing the device from
its housing)
2 frontplates (3) (for notes)
1 Ex approval (4) (only for Ex version devices)
1 Operating Instructions (5), in three languages: German,
French, English
SINEAX
B 811
SINEAX
B 811
ON
ON
(3)
(2)
(4)
4.96

3. Specification and ordering information

1. Mechanical design
Housing S17
)
Standard / 24... 60 V DC/AC
Standard / 85...230 V DC/AC
[EEx ia] IIC
MSK intrinsically safe
85...110 V DC/85...230 V AC
MSK intrinsically safe
3. Output signals
measuring outputs A1aund A12
0... 5 V, R
≥ 250 Ω
ext
1... 5 V, R
≥ 250 Ω
ext
0...10 V, R
≥ 500 Ω
ext
2...10 V, R
≥ 500 Ω
ext
Non-standard 0...> 5 to 0...15 V
Live zero > (1...5) to 3...15 V
0...20 mA, R
≤ 750 Ω (500 Ω)
ext
4...20 mA, R
≤ 750 Ω (500 Ω)
ext
Non-standard 0...1 to 0...< 20 mA
Live zero 0.2...1 to < (4...20) mA
4. FSK (Field communications protocol)
Not designed for communications protocol
Designed for FSK communication
at field output A12
Designed for FSK communication
at measuring output A1
5. Input circuit fault detection
(Open/short-circuit detection)
Open-circuit < 3.6 mA; short-circuit > 21 mA
Open-circuit 1 to 4 mA;
short-circuit 20 to 23 mA
6. Response to an input circuit
Output signal linear response
Increasing output signal >>>
Decreasing output signal <<<