Calcium Hardness; Ca5; Copper; Cu6 - Exact Micro 7+ Manuel D'instructions

Table des Matières
Calcium Hardness Test Procedure

CA5

REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: CA5
3
Press and re-press the MENU button until the display shows the parameter CA5.
FILL METER WITH SAMPLE
4
Rinse the CELL 2 or 3 times with the water sample you will be testing - rinsing minimizes
with the water sample.
ZERO METER
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
Strip Micro CA, Part No. 486629 into the CELL and immediately press READ.
®
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip after "1" on the display disappears. The cursor
will move across the display while the meter prepares to measure the sample. Record result
displayed (this result is automatically stored in CA5). After testing is completed, rinse cell
immediately and use brush to remove any residual from previous test.
CA5: Calcium reacts with Oxalic acid as it is released from the strip to form a white precipitate, directly
proportional to the Calcium concentration. Values are reported as Calcium Carbonate.
Copper Test Procedure

CU6

REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: CU6
3
Press and re-press the MENU button until the display shows the parameter CU6.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing
ZERO METER*
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
Strip Micro CU, Part No. 486632 into the CELL and immediately press READ.
®
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip after "1" on the display disappears*. The
display will immediately start counting up from 1 to 20 (this extra time allows more thorough color
development). At 20 sec, the cursor will move across the display while the meter prepares to
measure the sample. Record result displayed (this result is automatically stored in CU6). After
testing is completed, rinse cell immediately.
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
All manuals and user guides at all-guides.com
Strip Micro CA, Part No. 486629 from the bottle before
®
Strip Micro CU, Part No. 486632 from the bottle before
®