Bromine DPD-1 Test Procedure
CL
REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
SELECT TEST: BR3
3
Press and re-press the MENU button until the display shows the parameter BR3.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the water sample.
ZERO METER
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
Strip Micro CL (DPD-1), Part No. 486637 into the CELL and immediately press
®
READ. This starts the 20 SECOND countdown timer. During this time move the strip in a gentle
back and forth motion (approx. 2 strokes/sec). Remove and discard the strip after "1" on the
display disappears. The cursor will move across the display while the meter prepares to
measure the sample. Record result displayed (this result is stored in BR3 memory). After
testing is completed, rinse cell immediately and press MENU to select next test.
Total Alkalinity Test Procedure
AL
REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
SELECT TEST: AL4
3
Press and re-press the MENU
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the water sample.
ZERO METER*
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
Strip Micro AL, Part No. 486641 into the CELL and immediately press READ. This
®
starts the 20 SECOND countdown timer. During this time move the strip in a gentle back and forth
motion (approx. 2 strokes/sec). Remove and discard the strip after "1" on the display disappears*.
NOTE: For water temperatures above 95°F/35°C (hot tubs), remove and discard the strip when the
timer displays "10", countdown continues. For the hot water samples, a 10-second dip time is best.
The cursor will move across the display while the meter prepares to measure the sample. Record
immediately.
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.