DPD-4
(Total Chlorine or Ozone)
CL1
REMOVE STRIP
1
Remove one (1) eXact
486670 from the bottle before beginning the test. Set the strip in a dry, convenient place and
recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: CL1
3
Press and re-press the MENU button until the display shows the parameter CL1.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing
capacity (4ml) with the water sample.
ZERO METER*
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
Sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
Strip Micro CL (DPD-4), Part No. 486670 into the CELL and immediately press READ.
®
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back and forth
motion. Remove and discard the strip after "1" on the display disappears*. The cursor will move across
the display while the meter prepares to measure the sample. Record result displayed (this result is
automatically stored in CL1).
PRESS READ AGAIN
7
Press READ again and the meter will count down and display the next Total Chlorine result. If this
reading matches the previous result, then record this as Total Chlorine value (this result is stored
in CL1). After testing is completed, rinse cell immediately. Record the Total Chlorine as the highest
value the meter displayed.
NOTE: Standard Method (4500-Cl G, procedure for total chlorine) requires the reading to
be made after 2 minutes. 2 minute wait is not necessary for Ozone measurements.
PH2
REMOVE STRIP
1
Remove one (1) eXact
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: PH2
3
Press and re-press the MENU button until the display shows the parameter PH2.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the water sample.
ZERO METER*
5
Press the ZERO/ON button. When the display shows 0.0 PH, the sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
Strip Micro PH, Part No. 486639 into the CELL and immediately press READ. This
®
starts the 20 SECOND countdown timer. During this time move the strip in a gentle back and forth
motion (approx. 2 strokes/sec). Remove and discard the strip after "1" on the display
disappears*. The cursor will move across the display while the meter prepares to measure the
sample. Record result displayed (this result is automatically stored in PH2). After testing, rinse cell
immediately.
NOTE: For best results, Total Alkalinity of the sample should be 40-140 ppm.