b) Before using the sterilizer for sterilization, read the Instruction Manual
provided by the equipment manufacturer and follow the instructions.
c) Do not use hot air sterilization and radiation sterilization as this may
result in damage to the product;
d) Please use the recommended sterilization procedures for sterilization.
It is not recommended to sterilize with other sterilization procedures
such as ethylene oxide, formaldehyde and low temperature plasma
sterilization. The manufacturer assumes no responsibility for the
procedures that have not been recommended.
If you use the sterilization procedures that have not been recommended,
please adhere to related effective standards and verify the suitability and
effectiveness.
* Fractional pre-vacuum procedure = steam sterilization with repetitive
pre-vacuum. The procedure used here is to perform steam sterilization
through three pre-vacuums.
6.9 Storage
6.9.1 Store in a clean, dry, ventilated, non-corrosive atmosphere with a
relative humidity of 10% to 93%, an atmospheric pressure of 70KPa to
106KPa, and a temperature of -20 °C to +55 °C;
6.9.2 After sterilization,the product should be packaged in a medical
sterilization bag or a clean sealing container, and stored in a special
storage cabinet. The storage time should not exceed 7 days. If it is
exceeded, it should be reprocessed before use.
Notes: