Télécharger Imprimer la page

E-Mode: Set The Emissivity - Wohler IR Temp 310 Mode D'emploi

Thermomètre à infrarouge

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Fig. 7: warning when ambient tempera-
ture sensor is plugged in
Lo
Fig. 8: Display after alarm
12.2

E-Mode: Set the emissivity

Selecting the measurement mode
IRT (infrared temperature) – Tamb (temperature ambient)
> 6,5 °C:
Red side flashing
The bar shows red color
Hi is displayed beside the THB icon
IRT – Tamb < - 6,5 °C:
Blue side flashing
The bar shows blue color
Hi is displayed beside the THB icon
IRT – Tamb > 5,75 °C or < -5,75 °C:
or
The buzzer and optional vibration functions will
start at the same time.
According to the following value, "CHK" (check) is
displayed beside the THB mode icon. In this case
please continue to observe.
The bar colour changes according to the Temper-
ature:
4,25 °C < IRT – Tamb < 6,5 °C: red
- 3,5 °C ≤ IRT – Tamb < 4,25 °C: yellow
- 6,5 °C < IRT – Tamb < -3,5 °C: blue
Default emissivity: 0,95
The emissivity can be changed from 0.10 (10E) to
1 (100E).