MMD 540H Hybrid Multimeter | Push-button Functions | 11
Peak Hold
The Peak Hold function captures the peak
AC or DC voltage or current. The meter
can capture negative or positive peaks as
fast as 1 millisecond in duration.
1.
Set the function switch to the Amps or
Volts.
2.
Press and Hold the PEAK button until
"CAL" appears in the display. This procedure will zero the range
selected & the meter will go into manual ranging mode.
3.
Press the PEAK button, Pmax will display.
4.
The display will update each time a higher positive peak occurs.
5.
Press the PEAK button again, Pmin will display. The display will
now update and indicate the lowest negative peak.
6.
To return to normal operation, press and hold the PEAK button
until the Pmin or Pmax indicator switches off.
NOTE: If the Function switch position is changed after a
calibration the Peak Hold calibration must be repeated for the
new function selected.
MMD 540H Instruction Manual | Version 1 | February 2008