Electrical Connection; External Connections - Baumer CombiLyz AFI4 Notice D'utilisation

Masquer les pouces Voir aussi pour CombiLyz AFI4:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CombiLyz AFI4/AFI5
Inductive conductivity transmitter
6.

Electrical connection

6.1 External connections

9 A voltage supply of 15 to 35 V DC is provided.
Switch off supply voltage.
f
Connect the sensor in accordance with pin
f
assignment.
NOTICE
Damage to sealing or connector!
Tightening the knurl with too much force can damage the connector or the O-ring inside the cable gland.
Only tighten the knurl by hand with a maximum torque of 0.6 N.
f
Do not use any tools when tightening the knurl.
f
Left side (front view)
Connector type
M12, 4-pin
Right side (front view)
Connector type
M12, 4-pin
M12, 8-pin (with relay output)
Pin 2 in left connection and pin 2 (4-pin connection) or pin 7 (8-pin connection) in right connection can be
connected as a common minus for both conductivity and temperature signal output.
8 / 52
Conductivity measurement
Pin
1
2
3
4
Pin
1
2
3
4
1
2
3
4
5
6
7
8
Operating Instructions
M12, 4-pin
4
1
Color
Function
Brown
Supply +
White
Conductivity –
Blue
Supply –
Black
Conductivity +
Color
Function
Brown
S1
White
Temperature –
Blue
S2
Black
Temperature +
White
S1
Brown
Temperature +
Green
Relay 21
Yellow
Relay 22
Gray
Relay 11
Light red
Relay 12
Blue
Temperature –
Red
S2
M12, 8-pin
5
3
4
6
7
3
8
1
2
2
15 ... 35 V DC
4 ... 20 mA
15 ... 35 V DC
4 ... 20 mA
External input
4 ... 20 mA
External input
4 ... 20 mA
External input
4 ... 20 mA
4 ... 20 mA
External input
www.baumer.com