Electrical fast
±2 kV, 100 kHz for power supply line
transient/ burst
±1 kV, 100 kHz for input/output line
IEC61000-4-4
Surge
± 0.5, 1 kV (line to line)
IEC61000-4-5
± 0.5, 1, 2 kV (line to earth)
Voltage dips, short
Voltage Dips:
interruptions and
i) 100% reduction for 0.5 period,
voltage variations
ii) 100% reduction for 1 period,
on power supply
iii) 30% reduction for 25/30 period,
input lines
Voltage Interruptions:
IEC61000-4-11
100% reduction for 250/300 period
Power frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80 MHz
Conducted RF
6 Vrms in ISM bands
IEC 61000-4-6
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
±2 kV, 100
kHz for
power
supply line
± 0.5, 1 kV
(line to line)
230 Vac
30 A/m
30 A/m
3 Vrms
3 Vrms
0,15 MHz –
0,15 MHz – 80 MHz
80 MHz
6 Vrms in ISM and
6 Vrms in ISM
amateur radio
and amateur
bands between
radio bands
0,15 MHz and 80
between
MHz
0,15 MHz
80 % AM at 1 kHz
and 80 MHz
107
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of atypical
commercial or hospital
environment
Mains power quality
should be that of atypical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
this device requires
continued operation
during power mains
interruptions, it is
recommended that the
device be powered from
an uninterruptible power
supply or a battery.
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Portable and mobile RF
communications
equipment should be
used no closer to any part
of this device, including
cables, than there
commended separation
distance calculated from
the equation applicable to