Specifications
1-1 General Specifications
Dimensions : 82 mm (W) x 208 mm (L) x 41 mm (H)
Weight : 360 grams (including battery)
Power requirement : 9Vx1 battery
Auto power Off : The meter automatically turns
the power off after powering up and no operation
for 35 ± 10minutes.
Battery life (based on Alkaline Battery) : 45 hours
Maximum Conductor Size : 35mm
LVD : EN61010-1, EN61010-031, EN61010-2-032
EMC : EN61326-1
Installation Category :
CAT.III. 600V, CAT.II. 1000V for Jaw.
The circuits directly connected to Low-voltage
installation.
The building installation.
The source of the Low-voltage installation.
1-2 Environmental Conditions
Indoor Use.
Maximum operating altitude : 2000m (6562ft)
Operating temperature : 0°C ~ 30°C,
30°C ~ 40°C,
40°C ~ 50°C,
Storage temperature :
-20 to +60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Pollution Degree : 2
Shock vibration :
MIL-PRF-28800F for A class 2 Instrument
Drop Protection :
4 Feet Drop to hardwood on concrete Floor
Type of sensing : Hall effect sensing for AC and DC
8
80%RH
75%RH
45%RH