Calibration Mode

Examination of the Photometer

The verification standard kit for the P7 Advanced is designed to assure
the user of the accuracy and the reliability of the results related to the
integrated wave lengths.
Order code Verifikation-Standard-Kit: W3T330121
The kit contains one zero standard, 3 different vials for checking 3
different wave lengths.
To check the device with a Verifikation Standard Kit, proceed as follows:
Zero
Press the [ZERO/TEST] key and hold.
Test
On
Switch the unit on using the [ON/OFF] key.
Off
. . .
Allow the 3 decimal points to be displayed before releasing the [ZERO/
TEST] key.
: AbS
The instrument displays for a short time
followed by a measurable wavelength.
A530
Scroll through wavelengths using the [MODE] key.
Mode
Place the supplied blank ("Zero" Verifikation Standard) in the sample
chamber making sure that the marks are aligned
Place the supplied Verifikation Kit cover on the adapter.
Press the [ZERO/TEST] key.
Zero
Test
The symbol flashes for approx. 8 seconds.
AbS
The display shows:
0.0.0
WT.050.311.000.DM.IM.0119
Display
Wavelength
A530
530 nm
A560
560 nm
A610
610 nm
P7 Advanced
.
57