To Lock The Keypad - Nutricia Flocare Infinity PMS2655U Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

+
verrou (Locked)
d
pURgER
-
Non verrou (Unlocked)
Hold the '+' key down and press 'PURGER' for 2 seconds. The pump enters the set up mode. Press the
"PURGER" key to switch to the next setting:
"NON VERROU" the keypad (-) or "VERROU" the keypad (+). In the "VERROU" mode the feeding program
of the pump can no longer be changed.
Use the "+" or "-" key to change the setting from "verrou" to "non verrou" or vice versa.
+
BIp FIN dOSE (Beep high)
d
pURgER
-
NO BIp (Beep low)
Hold the '+' key down and press 'start/stop' for 2 seconds. The pump enters the set up mode. Press the
"PURGER" key to switch to the next setting:
"BIP FIN DOSE" (+): the pump will give an alarm when the dose is administered or "NO BIP" (-): the pump
will not give an alarm after administering the dose. Use the "+" or "-" key to change the setting from "BIP
FIN DOSE" to "NO BIP" or vice versa.
+
LUM ON
d
pURgER
-
LUM OFF
Hold the '+' key down and press 'start/stop' for 2 seconds. The pump enters the set up mode. Press the
"PURGER" key to switch to the next setting: "LUM ON" (+) the light of the LCD remains on when the pump
is connected to the mains. "LUM OFF" (-) the light of the LCD switches off after a few seconds even if con-
nected to the mains. Use the "+" or "-" key to change the setting from "LUM ON" to "LUM OFF" or vice versa.
+
OUTp pdMS
d
pURgER
-
OUTp NRSE
Hold the '+' key down and press 'start/stop' for 2 seconds. The pump enters the set up mode. Press the
"PURGER" key to switch to the next setting: "OUTP PDMS" (+) to connect with a Patient Data Monitoring
System. "OUTP NRSE" to connect with a NURSE CALL system. Use the "+" or "-" key to change to setting
from "OUTP PDMS", to "OUTP NRSE" or "OUTP OFF" and vice versa.
Press the
key and hold it down for 2 seconds to exit the SET UP mode. The settings are automatically
saved.
ZMNL900071_AW.indd 27

TO LOCK THE KEYpAd

TO MUTE WHEN dOSE dONE
TO SWITCH LIgHT ON
d
2 sec
TO CONNECT WITH NURSE CALL OR pdMS
d
2 sec
-25-
UK
11/7/11 4:35 PM

Publicité

Table des Matières
loading

Produits Connexes pour Nutricia Flocare Infinity PMS2655U

Table des Matières