Alarm Functions And Safety Features; Maintenance - Nutricia Flocare Infinity PMS2655U Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

• In case of battery failure, caused by, for example, an excessive temperature condition (which may resolve itself
with time) or by a complete battery failure, the battery symbol will flash. Turn the pump "off" and contact your
healthcare provider or refer the pump to service.
The power connector at the side of the pump offers the option of connecting the pump to an external alarm
system or Patient Data Monitoring System (PDMS).
• The Flocare
Infinity™ Nurse Call (corporate code 35752) enables connection to an external alarm
®
system. This external alarm system may be with open or closed contact. Follow the instructions for use of
the Flocare
Infinity™ Nurse Call to connect the pump.
®
• The Flocare
®
Infinity™ PDMS Cable (corporate code 35776) enables connection to an external PDMS
system. Follow the instructions for use of the Flocare

ALARM FUNCTIONS ANd SAFETY FEATURES

In case any of the problems listed on the problem solver chart occur, the pump delivers an audible and visual
alarm and stops working. The back light of the LCD automatically switches on.
Exception to this is the low battery alarm "BATT", in this situation the pump continues working.
Action in the event of an alarm:
Check the type of alarm displayed by the large display characters.
• Press the "START/STOP" key to stop the audible and visual alarm.
• Correct the cause of the alarm as described in the table.
• Start the pump again by pressing "START/STOP".
The problem solver chart on the next page gives a clear explanation of the alarms.

MAINTENANCE

Cleaning:
• Always unplug the pump prior to cleaning to avoid electric shock hazard.
• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor) with warm
soapy water, a 5% bleach solution in water, a multipurpose disinfectant cleaner or a soft cloth.
• The Flocare
Infinity
®
merge the pump!
• Always maintain the rollers on the rotor in a clean state to ensure they spin freely.
• The adaptor normally does not require cleaning. When desired, a dry or slightly damp cloth may be used
to clean the outside surface of the adaptor. Make sure the adaptor is disconnected from the wall outlet.
Safety notes:
• The manufacturer recommends an inspection of the pump at an authorised service centre every 2 years.
• If any fault occurs during use, or if the pump is dropped, it should be checked by authorised technical per-
sonnel prior to use.
• Replace the Flocare
growth of harmful bacteria. Dispose of Flocare
• Do not use the Infinity™ alarm systems to trigger actions related to secondary (electrical) medical devices
(e.g. a volumetric or syringe pump).
• Do not use pump functions (e.g. the fill set function) for any other purpose than described in this manual,
as this may cause the pump to incorrectly calibrate.
ZMNL900071_AW.indd 24
UK
CONNECTION TO AN ExTERNAL ALARM OR pdMS SYSTEM
pump may be rinsed by holding under a stream of warm, clean water. Do not sub-
Infinity
pump set every 24 hours to maintain delivery accuracy and prevent the
®
®
®
Infinity™ PDMS Cable to connect the pump.
Infinity
disposable sets properly, as required by local law.