Administration set is blocked or
kinked between the container and
the pump.
NO FLOW IN
Administration set or feeding tube
is blocked or kinked after the
pump, at the patient side.
NO FLOW OUT
The battery is discharging and is
running low.
LOW BATTERY
The battery is almost fully dis-
charged and should be plugged in
immediately; otherwise Compat
LOW BATTERY
Ella
will shut down shortly.
®
Battery is completely depleted
and must charge briefly before
pump can be used. Pump will
display
READY TO USE
CHARGE BATTERY
the pump can be restarted.
The pump is in pause mode and 2
minutes have passed without any
pump activity.
PUMP WAITING
Check for any kinking or blockage in the administration set tubing between the
tube feed container and the pump.
If necessary, reposition the administration set, pump or nutrition container to ensure
feed can flow freely.
Open the door, check the administration set installation, and re-close the door.
Restart feeding.
Check for any kinking or blockage in the administration set tubing between the pump
and the feeding tube.
If necessary, reposition the administration set or pump to ensure feed can flow freely.
Check that the feeding tube is clear and not kinked. Flush feeding tube if necessary.
Open the door, check the administration set installation, and re-close the door.
Restart feeding.
Connect pump AC power via pump AC adapter (see page 4).
Charge battery.
Acknowledge the
READY TO USE
when
Resume administration or switch pump off.
Extend inactivity timeout by another 2 minutes by pressing
message by pressing
.
.
EN
57