RayTek MI3 Serie Manuel D'utilisation page 103

Table des Matières

Publicité

Start
Size
Modbus Access Data
address
[bits]
80
32
input register
90
16
holding register short
100
8
discretes input
110
8
discretes input
120
16
holding register short
130
16
holding register short
410
32
input register
420
32
input register
430
16
input register
510
16
holding register short
520
16
holding register short
5<k>1
16
holding register short
5<k>2
16
holding register short
5<k>3
32
holding register float
5<k>5
32
holding register float
5<k>7
32
holding register float
MI3
Content
Type
float
Box Temperature
Switch panel lock
bit field Get connected heads
bit field Get registered heads
Laser control (only with
laser)
Relay alarm output
control
float
analog input 1 value
float
analog input 2 value
short
Trigger
analog output 1 mode
analog output 2 mode
Analog output k source head number or 0:
Analog output k source
parameter
Analog output k fixed
temp. value
Analog output k bottom
temp. value
Analog output k top
temp. value
Rev. G3 Oct 2020
Values
0: unlocked, 1: locked
bit 0: head 1 ..
bit 7: head 8
bit high: head connected
bit low: head disconnected
bit 0: head 1 ..
bit 7: head 8
bit high: head registered
bit low: head not registered
0: off, 1: on
0: off
1: on
2: norm. open
3: norm. closed
0 .. 5 V
0 .. 5 V
0: off
1: on
5: TCJ
6: TCK
7: TCR
8: TCS
9: 0...5 V
10: 0...10 V
99: disable (tristate)
0 – 0...20 mA
4 – 4...20 mA
9 – 0...5 V
10 – 0...10 V
99 – disable (tristate)
fixed value from 5<k>3
1: internal temp. of 5<k>1
2: object temp. of 5<k>1
value within range set in address
5<k>0
device bottom temp. .. device top temp. L<k>O
device bottom temp. .. device top temp. H<k>O
Modbus
MI3[M]
command
XJ
J
HC
HCR
XL
KB
TV1I
TV2I
XT
XO1O
XO2O
O<k>O
O<k>O
-
103

Publicité

Table des Matières
loading

Produits Connexes pour RayTek MI3 Serie

Ce manuel est également adapté pour:

Mi3commMi3mcommnMi3mcomm

Table des Matières