6.
Reinsert a new single-use trapper. Check the trapper is conform (presence of no-return membrane).
7.
Take a single-use pipette and check there are no defects or foreign bodies.
Insert the pipette into the trapper and check it fits tightly into the trapper. The device is ready for a new measurement. Remind the patient not to inhale in
the pipette.

3.4.2.Mode Monitoring

Description
This mode records the level of atmospheric carbon monoxide over a time defined in the TABATABA
Monitoring can be run for a recording time of:
1/4 hour
1 hour
8 hours
24 hours
In all cases, a measurement is taken every 1 minute. After each recording, the TABATABA
TWA (Time Weighted Average)
STEL (Short Term Exposure Limit)
Maximum value measured (C)
Use
1.
Do not use CO trapper or pipette.
2.
Place the TABATABA® V2 cursor on Monitoring mode (M) and switch on via the On/Off button.
3.
While pressing on the Unit/0 button, press the On/Off button.
The "Init" message appears, and recording begins.
4.
5.
Press the On/Off button again and the TABATABA
As soon as recording begins, the TABATABA
displays "RUN" and the measured value, then goes on standby. The TABATABA
®
the TABATABA
V2 successively displays the corresponding TWA, STEL and C values. While pressing on the Unit/0 button, press the On/Off button to
run a new monitoring.
®
The TABATABA
V2 can store two monitorings, the TWA, STEL and C values displayed corresponding to the values of the last monitoring recorded.
If the storage memory is full, the "Full" message is displayed for a few seconds. A long press on the Unit/0 button deletes all monitorings. The "CLR"
message appears, indicating that the memory has been deleted.
A monitoring in progress can be stopped by activating the device (On/Off button). The screen displays "RUN".
Validate by pressing the OK/SETUP button.

3.4.3.Alarm Mode

Description
In this mode, a sound and visual alarm is set off when the preset threshold (during set-up of the device) has been reached.
®
The TABATABA
V2 has two alarm thresholds:
Low threshold alarm
High threshold alarm
When an alarm threshold is reached, the device sets off an intermittent sound and visual alarm (red LED). The alarm signal remains active for at least 10
seconds.
These two thresholds are differentiated by a more rapid frequency signal for the high level.
Use
1.
Do not use CO trapper or pipette.
2.
Place the TABATABA® V2 cursor to Alarm mode.
Press the On/Off button. The TABATABA® V2 displays "ALAR". It then goes to low consumption mode (displayer blanks).
3.
4.
A measurement is taken every 20 seconds. A short flash is emitted by the green LED if the batteries have sufficient power.
Another press on the On/Off button reactivates the device, which displays "Init" then takes an instantaneous measurement and displays it for
5.
20 seconds. The device then returns to low consumption.
Ce document est la propriété de la Société FIM MEDICAL. Il ne peut être reproduit ou communiqué sans son autorisation.
This document is the property of FIM MEDICAL. Its contents cannot be reproduced or divulged without the company's approval.
V2 displays "RUN".
®
®
V2 performs an initial measurement, then switches off to save the batteries. Every minute, the screen
®
V2 set-up.
®
V2 displays the following values:
®
V2 can store two monitorings. If a monitoring has already been recorded,
V02.07.00
Août/August 2021
19