8. Status- and Error-Codes
8.1 Logger Display
CAL
Calibration active
Strt
The logger is waiting for the recording start, according to the settings in the datalogger setup menu.
StOP
Stop is reached if the logger is in the Start/Stop mode and the memory is full. Logging has been stopped. For a
new logging period, the logger has to be reprogrammed. This is the delivery status of the PT100 loggers TP2 and
TP4-L.
Ld.
The calibration parameters of the rH/T sensor are automatically uploaded when the sensor is connected to the
logger.
Prn
Data are printed
b.F.
Battery voltage is too low
CAL.E. Error during calibration
C.F.
Datalogger is faulty
Err
Error during printing
HHHH Logger keypad is defective
L.C.
Datalogger is faulty
Ld.F.
Error during upload of rH/T sensor parameters
n.c.
No sensor is connected
S.C.
Sensor has a short circuit
O.F.
Over flow of measurement value
U.F.
Under flow of measurement value
8.2 Logger Status in elproLOG ANALYZE
RAM IMG-BMP destroyed
This error message is visible in the logger status information, line: Module time. The reason for such an error message
might be, a battery replacement where the battery change time has not been set. (see also chapter 2.7 Maintenance).
9. Hints
TH1 & TH2
For sensor exchange, the corresponding humidity channel (Channel 2 or 4) has to show n.c. on the
display at least twice before the new sensor is connected to the logger! Please take care about the
sensor position before connecting it. There is no force on the sensor needed to plug it on. See also the
printing on the casing of the logger (TH1 only).The maximum cable length for probe 1 and probe 2
together should not exceed 20m. (Example: 1x3215-S20 oder 2x3215-S10)
TN2
In case of an alarm no entries are made into the alarm log.
Every time a data exchange with elproLOG ANALYZE software happens the function of the display will
be tested.
TN3-P
Every time a data exchange with elproLOG ANALYZE software happens the function of the display will
be tested.
Digital Inputs
In some of the documents the digital inputs are labelled with DigitIn. The same digital inputs are
labelled in the elproLOG ANALYZE software with D only.
Software
ECOLOG data loggers require an evaluation software of version 3.20 or higher.
CD-ROM
For more product information please see elproLOG ANALYZE CD-ROM.
D-EZ-6003Bh
Getting Started
ECOLOG
Seite - 23