Specifications - Eurotron IRtec P IVT Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

All manuals and user guides at all-guides.com
Distance & Spot Size
As the distance (D) from the object increases, the spot size (S) of the
area measured by the unit becomes larger. The relationship between
distance and spot size for each unit is listed below. The focal point for
each unit is 914mm (36"). The spot size indicates 90% encircled
energy.

3. Specifications

IR temperature measurement
Temperature range
D:S
Accuracy
Display resolution
Repeatability
Response time
Spectral response
Emissivity
Manual NTA70006-700A – IT/EN/FR
-50°C to 1000°C
50:1
 1%  1.0°C (1.8°F) from 20 to 500°C (68
to 932°F)
 1.5% from 500 to 1000°C (932 to 1832°F)
 3.5°C from -50 to 20°C (-58 to 68°F)
0.1°C (0.1°F) <1000°C
1°C (1°F) >1000°C
 1.5°C (2.7°F) from -50 to 20°C (-58 to
68°F)
 0.5% or 0.5°C (0.9°F) from 20 to 1000°C
(68 to 1832°F)
150 ms
8 – 14 µm
Digitally adjustable from 0.10 to 1.00
24

Publicité

Table des Matières
loading

Produits Connexes pour Eurotron IRtec P IVT

Table des Matières