or isopropyl alcohol and gently rub the surface, until no more residue is visible.
„ Replacing the Electrode
The TRACER is shipped with an electrode attached. Electrode life is limited and
is dependant on the frequency of use and care. If the electrode needs to be
replaced, follow these steps for removing and connecting electrodes.
1. Turn meter off.
2. To remove the electrode, turn the collar counter-clockwise and remove it.
3. Gently rock the electrode from side to side, pulling it downwards, until it
disconnects from the meter.
4. To attach an electrode, carefully plug the electrode into the meter socket.
Note that the electrode connector is keyed to ensure a proper connection.
CAUTION: Take case to align pins carefully. Bent or broken ins will cause the
meter to malfunction.
5. Tighten the electrode collar firmly enough to make a good seal. A rubber
gasket will seal the electrode with the meter.
TROUBLESHOOTING
Power on
Batteries
but no display
Batteries
Batteries
Unstable
Electrode
readings
Electrode
Electrode
Electrode
Electrode
Slow response
Clogged junction
time or reading
drift
Strong alkaline
measurement
Deteriorated gel layer
Protein coating on
electrode surface
Oil, paint, dyes,
suspended solids on
electrode
Dry bulb
Dehydrated membrane,
long term storage
without wetting
Static charge
Wiping electrode
Insert batteries
Verify correct polarity
Replace batteries
Immerse electrode more deeply in sample
Condition electrode before first use
Remove air bubbles caught under
electrode
Clean electrode
Replace electrode
Soak in 4.07M Potassium Chloride (KCl) at
60°C for 30 minutes
Soak in 0.1M Hydrochloric acid (HCl)
overnight
Replace electrode
Soak in 1g Pepsin dissolved in 10 mL of
0.1M HCl for 30 minutes or as needed
Rinse electrode alternately with materials
solvent then a buffer 7.00
Soak electrode tip in wetting cap filled
with 1 mL 4.0 buffer for 24-48 hours
Rinse electrode in 4.0 buffer and blot. Do
not wipe electrode
8