Using the Tango
Start therapy
Press
LCD.
If ramp is set, the ramp time displays on the LCD. When ramping is finished, therapy
continues at the prescribed pressure.
If no ramp is set, therapy starts immediately at the prescribed pressure.
Stop therapy
Press
Set ramp time
1.
2.
The ramp settings are (in minutes): Off, 0, 20, 5.
Set altitude
1.
2.
3.
The altitude settings are:
 (0–2000'/0–609 m),
2 (200–000'/60–29 m),
3 (00–6000'/220–828 m),
 (600–8500'/829–259 m).
Cleaning and Maintenance
WARNING
Beware of electric shock. Do not immerse the flow generator or power cord in water.
Always unplug the flow generator before cleaning and make sure it is dry before
reconnecting.
CAUTION
Do not hang the air tubing in direct sunlight as it may harden over time and
eventually crack.
Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing
or antibacterial soaps or scented oils to clean the air tubing or the Tango. These
solutions may cause hardening and reduce the life of the product.
Daily: Disconnect the air tubing and hang it in a clean, dry place until next use.
Weekly:
1.
Remove the air tubing and wash it in warm water and mild detergent. Rinse
thoroughly, hang and allow to dry.
2.
Before next use, reassemble the mask and headgear.
3.
Reconnect the air tubing.
Monthly:
Clean the exterior of the Tango with a damp cloth.
Check the air filter for holes and blockage by dirt. Replace it every six months, or more
often in a dusty environment.
For information on cleaning your mask or humidifier, refer to the manual supplied with
your mask or humidifier.
when the Tango is in standby mode. A circulating dash symbol displays on the
.
Repeatedly press
until the LCD displays the desired time.
Press
to accept changes and start therapy. (The Tango returns to standby mode
after 5 seconds if no button is pressed.)
Hold
down for 3 seconds until the current altitude setting is displayed.
Press
or
to change the altitude setting.
Press
to accept changes and return to standby mode. (The Tango returns to
standby mode after 0 seconds if no button is pressed.)