Guidance and manufacturer's declaration—electromagnetic immunity
The C-Series Tango system is intended for use in the electromagnetic environment specified below. The
customer or the user of the C-Series Tango system should assure that the device is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 6000--2
Electrical fast
transient/burst
IEC 6000--
Surge
IEC 6000--5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines.
IEC 6000--
Power frequency
(50/60 Hz)
magnetic field
IEC 6000--8
Note: Ut is the AC mains voltage prior to application of the test level.
IEC60601-1-2 test level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
± kV for input/output
lines
± kV differential mode
±2 kV common mode
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
0% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95%dip in Ut)
for 5 sec
3 A/m
Compliance level
±6 kV contact
±8 kV air
±2 kV
Not Applicable
± kV differential mode
±2 kV common mode
<2V
(>95% dip in 20V)
for 0.5 cycle
96V
(60% dip in 20V)
for 5 cycles
68V
(30% dip in 20V)
for 25 cycles
<2V
(>95%dip in 20V)
for 5 sec
3 A/m
0
Electromagnetic
environment—
guidance
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the C-
Series Tango system
requires continued
operation during power
mains interruptions, it is
recommended that the
C-Series Tango system
be powered from an
uninterruptible power
source.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.

Publicité