EN. Domistar, Hospistar, Neostar, PF4 - High Pressure regulators. Instruction for Use.
Appendix Nr 2
Scope
This appendix provides pressure outlet connection and disconnection procedure for
the owner and the end user. It is part of the Instruction book FOR Combivalves and
regulators.
AFNOR pressure outlets
Design conditions
Attached standard
Maximum working pressure
Above pictures represent an Oxygen AFNOR quick coupler and are given for infor-
mation for identification purpose. Quick coupler fitted on the product may be
slightly different special if it related to other gas. Connection / Disconnection as
well as service recommendations procedures remain the same.
Connection/Disconnection procedure:
NF S EN 90-116
6 bar
Front view
Connection:
1. Insert the male probe internal tube in the coupler
centre hole – position the male probe external lugs in
front of the coupler notch and push forward until it
stop.
2. Maintain the push effort and turn right until the male
probe external lugs are positioned in front of
the coupler round groove.
3. Release the push effort and ensure the male probe lugs
are engage.