Technical Specifi Cations; Electromagnetic Compatibility; (Emc) - DTF MEDICAL kitett MINIKIT DUO Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10
Technical specifi cations
Essential performance features:
Provide adequate suction so that
mothers can express their milk.
Control the rate of suction to best
meet the baby's sucking rhythm.
Equipment
with
compression
and
low
Intermittent suction.
Part applied to the body:
Breast shield.
NOTE:
The
CE
declaration
conformity and a technical description
can be supplied on demand.
Identifi cation:
Each
breast
pump
has
identifi cation that indicates the year
of manufacture.
11
Electromagnetic compatibility (EMC)
Electromedical devices require special precautions with regard to EMC and must be installed and set to operate in accordance
with the EMC information provided in this manual.
WARNING:
Avoid using this unit near or stacked onto other equipment as this may cause a malfunction. If this is necessary,
this appliance and the other appliances should be monitored to check they are operating normally.
WARNING:
Use of accessories, transducers and cables other than those specifi ed or supplied by the manufacturer of this
device may cause an increase in electromagnetic emissions or a decrease in the immunity of this device and cause improper
functioning.
WARNING:
Portable RF communication devices (including peripherals such as aerial cables and external aerials) should
not be within 12 inches (30 cm) of any part of the MINIKIT DUO breast pump, including cables specifi ed by the manufacturer.
Otherwise, performance of these devices may be impaired.
WARNING:
If the device malfunctions due to electromagnetic disturbances, the performance of the device may be
impaired and result in delayed patient care.
Wireless communication devices such as wireless home networking devices, mobile phones, cordless phones and their bases,
walkie-talkies can a ect the device, and it is recommended to keep them at a distance of at least 0.5 m from the device.
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
EMISSION TEST
RF emissions – CISPR 11
RF emissions – CISPR 11
Harmonic emissions (IEC 61000-3-2)
Voltage fl uctuations/Flicker (IEC 61000-3-3)
Device class (in accordance with Directive 93/42/EEC
amended by Directive 2007/47/EC)
Class of electrical safety device
moderate
fl ow.
Expected life of accessories
Capacity of the collection container
of
a
batch
Battery
Room temperature (T)
Room temperature (T)
for transport and storage
Relative humidity (H) for transport,
storage and operation
Atmospheric pressure (P) for
transport, storage and operation
Type of applied parts
Degree of protection
Expected life of the device
Weight of the device
Dimensions
Voltage
Length of the cable
Absorbed power
Suction Values
Rate
Sound level
Power supply
for operation
COMPLIANCE
Group 1
Class B
Class A
Compliant
IIa
Class II
BF
IP22
1 year
300 hours
100 ml
200 g without battery
120 mm x 90 mm x 60 mm
Input: 100-240V (AC) – 50/60Hz – 0.5A Max.
Output: 6V (DC) – 1.0A
1.8m
5.2 W
Stimulation phase: -80 to 140 mbar
Expression phase: -155 to 280 mbar
Maximum suction: <-330 mbar
40 to 120 cycles/min.
< 56 dB(A)
Alternating current 100-240 V - 50/60 Hz
4 x 1.5V - AA, LR06
+5°C < T < +40°C
-25°C < T < +70°C
15% < H < 90%
700 hPa < P < 1,060 hPa
EN
23