Technical Data
HandyStep
S with PD-Tips from BRAND, 20 °C 'Ex',
®
Nominal
Volume
volume
range (µl)
PD-Tip
0.1 ml
2 -
0.5 ml
10 -
1.0 ml
20 - 100
1.25 ml
25 - 125
2.5 ml
50 - 250
5.0 ml
100 - 500
10.0 ml
200 - 1000
12.5 ml
250 - 1250
25.0 ml
500 - 2500
50.0 ml
1000 - 5000
A* = Accuracy, CV* = Coefficient of variation
Error limits refer to the partial volume set relative to the PD-Tip, obtained when instrument,
PD-Tip and distilled water are equilibrated at ambient temperature (20 °C/68 °F) and with
smooth operation. The testing is according to DIN EN ISO 8655-5.
Important!
The HandyStep
®
manufacturers. The user must check their compatibility and perform a
function and volume test before use!
Stroke setting
of nominal volume
1
2%
=
10
8.0
50
4.0
4.0
4.0
3.5
2.5
1.5
1.5
1.5
1.5
S can also be operated with compatible tips from other
Accuracy Table
A*
± %
<
–
%
=
3
6%
5
10%
=
=
2.7
1.6
1.33
0.8
1.33
0.8
1.33
0.8
1.17
0.7
0.83
0.5
0.5
0.4
0.5
0.3
0.5
0.3
0.5
0.3
CV*
%
<
–
Stroke setting
=
of nominal volume
1
2%
3
6%
5
=
=
=
5.0
3.0
2.0
1.4
0.73
0.6
1.0
0.38
0.4
0.8
0.38
0.3
0.8
0.3
0.2
0.6
0.27
0.2
0.5
0.23
0.2
0.3
0.23
0.2
0.4
0.23
0.2
0.4
0.23
0.15
%
10%
33