Use the untreated ear if prescription ear drops or other ear medications have been placed
in the ear canal.
4.
Holding the thermometer for too long in the hand before taking a measurement can cause the device
to warm up. This means the measurement could be incorrect.
5.
Patients and the thermometer should stay in a room with a steady environment for at least 30 minutes.
6.
Before placing the thermometer sensor onto the forehead, remove dirt, hair, or sweat from the fore-
head area. Wait 10 minutes after cleaning before taking measurement.
7.
Use an alcohol swab to carefully clean the sensor and wait for 5 minutes before taking a measurement
on another patient. Wiping the forehead with a warm or cool cloth may impact your reading. It is ad-
vised to wait 10 minutes before taking a reading.
8.
In the following situations it is recommended that 3-5 measurements in the same location be taken and
the highest one taken as the reading:
New-born infants in the first 100 days.
Children under three years of age with a compromised immune system and for whom the
presence or absence of fever is critical.
When the user is learning how to use the thermometer for the first time until he/she has
familiarized himself/herself with the instrument and obtains consistent readings.
CARE AND CLEANING
Use an alcohol swab or cotton swab moistened with 70% alcohol to clean the thermometer casing and the
measuring probe. After the alcohol has completely dried out, you can take a new measurement.
Ensure that no liquid enters the interior of the thermometer. Never use abrasive cleaning agents, thinners or
benzene for cleaning and never immerse the instrument in water or other cleaning liquids. Take care not to
scratch the surface of the LCD screen.
ERROR AND TROUBLESHOOTING
Symptom
Failed to power on.
The reading is too
high.
Possible Cause
The battery level is too low.
Polarities of the batteries are re-
versed.
The thermometer is damaged.
The lens of the probe is dirty.
The distance of the item and target
is too far.
You have just come from a cold
environment.
The ambient temperature is not in
operable range.
Memory error or calibration is not
finished. Item will be turned off
automatically after 3 seconds.
In ear or forehead mode, T > 109.2°F
(42.9°C). In object mode, T > 199°F
(100°C).
Description & Solution
Replace with a new battery.
Ensure the batteries are in the right position.
Contact Escali.
Clean the lens with a cotton swab.
Keep in contact with forehead, or put the probe
into the Ear Canal.
Stay in a warmer room for at least 30 minutes
before taking a reading.
One long beep, and red backlit for 3 seconds.
Take a measurement under an ambient temper-
ature between 50.0°F (10°C) and 104°F (40°C).
One long beep, and red backlit for 3 seconds.
Contact Escali.
One long beep, the screen will be displayed in
red.