If you do not already have serial terminal software, free, open-source options are available online (i.e.,
RealTerm, etc.). You can download RealTerm* software from:
http://realterm.sourceforge.net/
After installation, click the Port tab and set the following:
Baud: 115200
Port: The appropriate VCP for your ProFLOW 6000 flowmeter
Parity: None
Data Bits: 8 bits
Stop Bits: 1 bit
HardwareFlow Control: None
Go to the USB menu entry in the ProFLOW 6000 flowmeter and turn the transmission on. Finally, click Open
on the Port tab in RealTerm and you will see the serial data stream begin in the terminal window (Figure 8).
If you would like to log the flow data, this can be done by clicking on the Capture tab. Set File to the name
and location of the log file that you would like to save and click either the Start: Overwrite or Start: Append
buttons appropriately (Figure 9).
*This software is not supplied or supported by Restek. User assumes all responsibility for the downloading and use of the program.
Figure 9: Screen capture of data collection.