Operating Instructions BA ULTIMATE
65 56 083 D3560.201.10.03.02
6

Follow-up

6.1

After each treatment session

NOTICE! Condition immediately, or at the latest, one hour after
Immediately after treatment
treatment.
✔ The bur instrument must not be moving.
Turbine
✔ Wear appropriate protective clothing.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Remove the bur instrument with tweezers.
3. Pre-disinfect directly at the treatment center [ → 23].
4. Remove the turbine.
5. Transport the turbine to the hygiene room in a suitable transport
container.
6 Follow-up
21