Guidance and manufacturer's declaration - electromagnetic immunity
The Volumed µVP7000 Series is intended for use in the electromagnetic environment specified below. The customer or the user of the Volumed
µVP7000 Series should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 610004-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
GA-7000-F 0707.doc
IEC 60601
Compliance level
test level
± 8 kV contact
± 8 kV contact
± 15 kV air
± 15 kV air
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for
± 1 kV for
input/output lines
input/output lines
± 1 kV differential
± 1 kV differential
mode
mode
± 2 kV common
± 2 kV common
mode
mode
<5 % U
<5 % U
T
T
(>95 % dip in U
)
(>95 % dip in U
T
for 0.5 cycle
for 0.5 cycle
40 % UT
40 % UT
(60 % dip in UT)
(60 % dip in UT)
for 5 cycles
for 5 cycles
70 % UT
70 % UT
(30 % dip in UT)
(30 % dip in UT)
for 25 cycles
for 25 cycles
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 5 sec
for 5 sec
400 A/m
400 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of typical commercial or hospital
environment.
Mains power quality should be that of typical commercial or hospital
environment.
As the Volumed µVP7000 Series is powered by an internal battery, no
special precautions must be taken.
)
T
31/31